Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 1.916 | 1.837 |
500.00 | 1.594 | 1.515 |
250.00 | 1.263 | 1.184 |
125.00 | 0.829 | 0.750 |
62.50 | 0.573 | 0.494 |
31.25 | 0.305 | 0.226 |
15.63 | 0.195 | 0.116 |
0.00 | 0.079 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 84-98% | 91% |
EDTA plasma(n=5) | 89-103% | 96% |
Heparin plasma(n=5) | 82-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 93-102% | 78-92% | 87-98% | 80-93% |
EDTA plasma(n=5) | 98-105% | 86-92% | 87-93% | 95-103% |
Heparin plasma(n=5) | 96-101% | 85-92% | 79-96% | 89-101% |